drive slow park latch does not work
Symptom
drive slow park latch does not work
Error message
solenoid fails 1) The Park Latch solenoid actuator falls off behind the LVLE PSU,this causes the latch to not be released when tube raised from park. The actuator can be put back in place and secured with a tiny blob of locktite,NO parts required. Early sets have a solenoid with an internal thermal switch that can fail,,,they do not have the park sensor pcb 2173060 and the solenoid clicks every time they move the tube. There is a mod kit to change park sensor pcb The Park Sensor PCB 2173060 (Hall effect device operating a relay instead of plunger switch) has very bad solder joints and fails due to vibration This causes the set to go SLOW and the solenoid latch not operating as it thinks the tube is not parked
Applies to
- Product
- AMX4 AMX4+
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray R&F issues
- ADx-E Advantx-E console displayed a 6C00 Error (Digital Imaging Subsystem error - Error status received in IDC Prep Reply) when the first exposure of detector bad pixel calibration was attempted; cali…
- AEC Backup Time message
- New style ion chamber does not have a preamp
- Precision 500D Table Fluoro Carriage Front Bumper is not properly adjusted, allows excessive travel toward the front of the table.
- Precision 500D Aurora, dose readout does not update when an overhead Rad Tube Exposure is made, at any technic due to abnormally low mR/mAS from the Rad Overhead XRT.
- P500D Aurora with M4 Software (May 2006 Release), Dose Readout Correct on Console, Dose Printer Prints 0.0 due to a M4 Software Bug