GE Healthcare · CTLightSpeedVCT260002453260002376
LightSpeed 2.X (Plus, QX/i) - GRE
Symptom
Customer sees a Pop-up window indicating the DAS temperatures are low. This does not prevent scanning only a warning. Customer sees a Pop-up window indicating the DAS temperatures are low. This does not prevent scanning only a warning
Error message
Error: 260002453 Data Acquisition System Temperature Is Low. See long description for a sample error message. Thu Nov 20 06:27:02 2008 Error: 260002453 Host: DAS Process: DCB Firmware File: ctemptst.c Line: 681 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 06:27:03:550 Log Series: 31 Data Acquisition System Temperature Is Low. Error: 260002376 DAS Temperature is LOW. See long description for a sample error message. Thu Nov 20 06:27:03 2008 Error: 260002376 Host: DAS Process: DCB Firmware File: ctemptst.c Line: 697 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 06:27:03:825 Log Series: 32 DAS Temperature is LOW. Lowest temperature was seen is on Converter Board 30. Measured: 10.0 degrees C. Acceptable Range: 15.0 to 62.0 degrees C. Converter Board Bit Map: 0x000020000000. Bit Map Indicates Failed Board(s) for Multiple Board Errors. (Board Bit Map defines: BIT47 = CnvBrd 48 ... BIT0 = CnvBrd 1)
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 2.X (Plus, QX/i) - GRE always an SDAS 4 slice H2_P_S4 (SDAS4 and H2 gantry) Lightspeed 2.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 3.X and 4.X. Always has an SDAS 4 slice Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2137958-4 SDAS Convertor Board (replaces 2137958-3) May use 2137958-3U converter Board if no current stock of the 2137958-4 See long description for details Part #: 2137958-4 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: -39999.60 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: CONVERTOR WITH BLACK BEUTY CONNECTOR FRU: Yes Status: Orderable Comments: OBSOLETE PART case 10788898 This part has been identified as obsolete and no short-term solution has been identified FE must identify and order a higher or lower level spare part as required Contact your support engineer on-line engineer or Parts ID team should you require support 212012173 06-26-07 REPLACES P/N 2137958-3 THE NEW -4 IS COMPATIBLE TO THE -3. MS20544 ON 3/7/03. ** FE ASSIST ** May use the USED part 2137958-3U converter Board if no current stock of the 2137958-4. (9/10/2007 212001261)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.