Collapsed Image and Gradient Link hammer Test Failures Due to the STIF or SRF TRF.

Symptom

MRIQ Collapsed Image Bright White Line Down or Across Middle of The Image

Error message

2253850 Gradient Link Hammer test failed. The GP received an invalid %c axis data value via the gradient data link. Check the gradient data and clock fiber optic cable connections 2253850 Gradient Link Hammer test failed. The GP received an invalid %c axis data value via the gradient data link. Check the gradient data and clock fiber optic cable connections

2244578 The GP is reporting a Gradient Amplifier % Axis Current Distortion Fault

2250082 Z Axis Gradient Data Overrange Occurred

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2280952-3 STIF Board See Long Description SRF TRF I/O or STIF 2280952 crosses up in FEMC to 2280952-3. Part Description is "STIF Assembly".
Area
MR