Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT19805171980509

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Prospective images slow to reconstruct. Some images came up and then a minute later the other images showed up.

Error message

Error: DriveReady SeekComplete Error See LD for more details Error: DriveReady SeekComplete. This was on a GRE3 console. System was having scan aborts with problems writing to SDDA. rsh darc cd /var/log tail –1000 messages | more (to view last 1000 lines of log) cat messages | grep Error (this will list the Errors only) Look for any Errors like the following which are DARC or VDARC disk drive errors: Aug 19 12:31:32 darc kernel: hda: dma_intr: status=0x51 { DriveReady SeekComplete Error } Aug 19 12:31:32 darc kernel: hda: dma_intr: error=0x40 { UncorrectableError } LBAsect=1980517, high=0, low= 1980517, sector=1980509 Aug 19 12:31:32 darc kernel: end_request: I/O error, dev hda, sector 1980509 Aug 19 12:31:32 darc kernel: hda: dma_intr: status=0x51 { DriveReady SeekComplete Error } Aug 19 12:31:32 darc kernel: hda: dma_intr: error=0x40 { UncorrectableError } LBAsect=1980517, high=0, low=1980517, sector=1980517

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5159833 VDARC - Volume Data Acq Recon Control Computer, 2nd Edition (Jarrell). See Long Description for Details.. Part #: 5159833 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: VDARC - Volume Data Acq Recon Control Computer, 2nd Edition FRU: Yes Status: Orderable Parts Offered as: Current : 5159833 Harvested : Not Available Exchange : Not Available Comments: Per Jarrell RSPL, 7/17/06, ms11892. ** FE Assist ** Must order software release 06MW29.7, p/n 5193989 and perform a LFC to support this FRU if system is not currently running 06MW29.7 or better.(12/8/06 212001261
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.