Pular para o conteúdo principal
GE Healthcare · MRSigna22493412249338108193495610639224936410599

System Intermittently Locks up with CM Init Failed Errors Due to APS.

Symptom

Scan Stops Intermittently and the Screen Locks Up

Error message

2249341 CM init failed. smNameFind failed. Communication to APS has failed.

2249338 CM init failed. msgQSmCreate failed when creating VxMP mailbox. Error seen with MGD Diags: 1081934956 4 1 Wed Apr 14 09:29:16 2004 2249338 2 CARD diagCoord cm_mp.c 306 Task ID: 0x3a652d0 Time: 10639 Object: APS CM init failed. msgQSmCreate failed when creating VxMP mailbox. errno = 0x3D0001.->

2249364 MGD Posix Sync Init could not find the shared data structure in the name database. 1081934956 3 1 Wed Apr 14 09:29:16 2004 2249364 4 CARD diagCoord mgdPosixSync.c 322 Task ID: 0x3f1fdf8 Time: 10599 Object: APS MGD Posix Sync Init could not find the shared data structure in the name database. vxWorks errno = 0x590001.

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2294300-18 APS2 Board This part is meant for use in Combined Asc Mgd (CAM) Chassis released in Q4 of 2007. It is supported by these software revisions or later: HDx 14.0 M5, HDe 14 M4 with Service Pack 4, and Ovation 5 (MFO) post-M4. MGD Chassis operating with 14.0 M5 or later may also be shipped with this part. MGD Chassis (NOT CAM) can substitute 2294300-16 if necessary. 212002950 02/14/2008.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues