Grainy Head Coil Images and RF Transmit Power faults due to Head TR Assembly.

Symptom

Head images with Quad Head coil are grainy, TG is high, RF power message. Error: 2225664 RF Amplifier fault 85. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. Check patient weight entered. amplitude: 15 duty cycle: 05 Fault in head amplitude. If you need to confirm the source of the fault, reset the power monitor and repeat the scan.

Error message

2225664 RF Amplifier fault 85 RF Transmit power has reached safety limit allowed for scan parameters and patient weight Error: 2225664 RF Amplifier fault 85. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. Check patient weight entered. amplitude: 15 duty cycle: 05 Fault in head amplitude. If you need to confirm the source of the fault, reset the power monitor and repeat the scan.

2224583 Auto Prescan Failed, Possible Data Acquisition Failure. File: APS:aps_cf Line: 15 Auto Prescan failed, possible data acquisition failure.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
46-264988G2 LPCC 1.5T HEAD T/R ASSEMBLY
Area
MR