Grainy Head Coil Images and RF Transmit Power faults due to Head TR Assembly.
Symptom
Head images with Quad Head coil are grainy, TG is high, RF power message. Error: 2225664 RF Amplifier fault 85. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. Check patient weight entered. amplitude: 15 duty cycle: 05 Fault in head amplitude. If you need to confirm the source of the fault, reset the power monitor and repeat the scan.
Error message
2225664 RF Amplifier fault 85 RF Transmit power has reached safety limit allowed for scan parameters and patient weight Error: 2225664 RF Amplifier fault 85. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. Check patient weight entered. amplitude: 15 duty cycle: 05 Fault in head amplitude. If you need to confirm the source of the fault, reset the power monitor and repeat the scan. 2224583 Auto Prescan Failed, Possible Data Acquisition Failure. File: APS:aps_cf Line: 15 Auto Prescan failed, possible data acquisition failure.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 46-264988G2 LPCC 1.5T HEAD T/R ASSEMBLY
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.