GE Healthcare · MRSigna22256432225607
Erbtec RF Amplifier Faults 16 and 57 Due to Doorknob Capacitors.
Symptom
Unable to scan with Erbtec errors
Error message
2225643 ERBTEC RF Amplifier fault 57 +3kV power supply unacceptable within 0.5 to 6 seconds after closing HV start relay 2225643 ERBTEC RF Amplifier fault 57. +3kV power supply unacceptable within 0.5 to 6 seconds after closing HV start relay. Possible defective HV rectifier/filter board, bad relay in AC Switching Module, or bad A/D on Processor Board. 2225607 ERBTEC RF Amplifier fault 16 HPC PORT ID Error: 2225607 File: PMH.m 1.29 Method: spi rf mon Line: 3035 ERBTEC RF Amplifier fault 16. 48V power supply unacceptable. Less than 42.9V or greater than 51.5V. Possible defective Power Control Board or Solid State Amplifier Board.
Applies to
- Models
- Signa
- Product
- Signa Advantage Includes 4.x and 5.x systems. The software that this solution is tied to is indicated by the Software concept below.
- Software
- 5.X
- Area
- MR