GE Healthcare · CTLightSpeedVCT1841515000423000002450016189000
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE
Symptom
Fast Calibration fails with many different errors. Fast Calibration fails with many different errors
Error message
Error: 184151 Scan Control timed out waiting for the Ready For Views message from the SRU See the long description for the full error code Sat Apr 17 20:29:44 2004 Error: 184151 Host: STC Process: Scan Control File: scTrigSeqActions.c 1.65 Line: 1365 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 50004/3/1 Type: Axial Exception Level: Pri/Most Severe Time: 20:29:43:355 Log Series: 676 Scan Control timed out waiting for the Ready For Views message from the SRU. This message is sent from the SRU once offset collection has completed. Possible causes: trigger failure, DIP failed to receive data, message lost, software error Error: 230000024 Timeout expired waiting for the DIP to receive a buffer of views. See LD for more details Error: 230000024 Timeout expired waiting for the DIP to receive a buffer of views. Wed Mar 22 03:38:05 2006 Error: 230000024 Host: darc Process: dataacq File: dip thread.cxx Line: 612 Function: Data Acquisition : Scan Data Acquisition Scan Type: Cine Scan: 50016/6/1 Scan Seq Id: 189000 Timeout expired waiting for the DIP to receive a buffer of views. Timeout value: 250 msecs Offsets collected: 404c598h bytes Minimum offsets: 1f544c4h bytes Active dat
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5117866-51 Host Computer, CTPET HP8200 with FX1500 Graphics Card and 4GB Memory - replaces -24, -28, and -48 versions See Long Discription for Details Part #: 5117866-51 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Host Computer, CTPET HP8200 with FX1500 Graphics Card and 4GB Memory FRU: Yes Status: Orderable Comments: Per ECO 2042338, 5117866-51 is a substitute for 5117866-28 for Brightspeed systems. 212008898, 6/13/07. ** FE Assist ** For VCT systems must have software release 07MW18.4, p/n 5213049 (deployed by FMI 25394, p/n 5195244), and perform a LFC to support this FRU if system is not currently running sw rls 07MW18.4. or newer. For other CT systems , must order sw releases 07MW11.10, p/n 5220443 (deployed by FMI 25389, p/n 5191771)and perform a LFC to support this FRU if system is not currently running the sw release, or newer (9/12/07 212001261) 5115333 VCT GOC5 HOST HP 8000 computer New part number 5117866-28 Description: Host Computer, CT USA VCT XW8200 Part #: 5117866-28 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Host Computer, CTPET HP8200 with FX1400 Graphics Card and 4GB Memory FRU: Yes Status: Orderable Comments: Added per VCT RSPL, 040522 ms13452 This part is direct replacement for 5115333 . ms13452 Part #: 5117866-24 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 60.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Host Computer, CT USA 4,8,16-slice XW8200 FRU: Yes Status: Orderable Comments: Added per pathfinder RSPL, this part is not replacement validated for xw8000. ms13452 *** 2ND OPINION NOTES *** This part has a High NDF ( No Defect Found ) rate. Attempt to reload OS and Application software before replacement. ( 212024490 - 12/20/2006 )
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required