Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT23000700423000701450029230007011

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Image artifact after fastcals failed with DAS/detector erors See long description for image Image artifact after fastcals failed with DAS/detector erors

Error message

Error: 230007004 Channel ###, Module ## Short to ground* FAILED X- AXIS EXCLUSION CRITERIA EXCLUSION CRITERIA *. Error: 230007004 can be used for any row, channel and module that is failing. Error: 230007004 Channel ###, Module ## Short to ground* FAILED X- AXIS EXCLUSION CRITERIA Examples: Sat Sep 17 08:08:02 2005 Error: 230007004 Host: amhvct Process: badchanmgr File: BadChannelMap.cxx Line: 584 Exclusion Test - Row 33:1B, Channel 900, Module 57 Short to ground* FAILED X-AXIS EXCLUSION CRITERIA *. Tue Aug 29 05:24:31 2006 Error: 230007004 Host: ct99 Process: badchanmgr File: BadChannelMap.cxx Line: 667 Exclusion Test - Row 1:32A, Channel 418, Module 27 Short to ground* FAILED X-AXIS EXCLUSION CRITERIA *.

Error: 230007014 Uncorrectable detector channels have been found for scan See LD for an example of entire error message. Fri Oct 28 09:32:03 2005 Error: 230007014 Host: presbyplano Process: badchanmgr File: badchanmgr.cxx Line: 1194 Uncorrectable detector channels have been found for scan 50029/2/2

Error: 230007011 Density Check - Module (X) exceeds number of shorts allowed (X) = modual number See LD for complete error Mon Oct 31 07:03:37 2005 Error: 230007011 Host: cpcvct2p Pr

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5134959 FRDM Module L-FINE 32 replaces 5123019 Used on VCT 32 Detector See Long Description for Details. 5134959 FRDM Module L-FINE 32 replaces 5123019-Used on VCT 32 Detector Part #: 5134959 Part Type: GP CONSUMABLES Part Weight: -98999.01 LBR Repairable: No Returnable: No Hazardous Material: N Description: L-FINE 32 Detector Module Kit FRU: Yes Status: Orderable Part #: 5123019 Part Type: MST DEFAULT LOAD Part Weight: 0.00 LBR Repairable: Not Available Returnable: Not Available Hazardous Material: N Description: L-FINE 32 DIGITAL MODULE ASSY MP3 FRU: Yes Status: Non Orderable
Area
CT

Full cause & step-by-step resolution