Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT290012003238395172290010005

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Recon self test failure on boot up; scanning would be allowed but no reconstruction of images. Recon self test failure on boot up; scanning would be allowed but no reconstruction of images

Error message

Error: 290012003 IG internal allocation error: 238395172 in CyclicMemBuffer.cpp, 213 Thu Apr 19 20:16:55 2007 Error: 290012003 Host: ig3 Process: image_generation File: IgTrace.cxx Line: 105 IG internal allocation error: 238395172 in CyclicMemBuffer.cpp, 213

Error: 290010005 Socket error: socket is terminated because IG dies (error text varies) See Long Description for error examples. Tue Mar 16 10:40:09 2004 290010005 darc data_restore request_object.cxx 503 Socket error: socket is terminated because IG dies Fri Nov 16 19:22:57 2007 Error: 290010005 Host: darc Process: data_restore File: request_object.cxx Line: 556 Socket error: socket event from IG

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.