GE Healthcare · MRSigna22255542225553
Guide pins are loose and damaged making it difficult to maintain a good connection.
Symptom
Can't get coil cable to stay connected
Error message
2225554 Multi-Coil T/R Driver Board Fault. An Open Circuit was Detected on Port 4. Thu Aug 21 12:06:10 1997 Suite: MRS1 Host: OC0 Proc: HPC_PORT_ID Error: 2225554 File: PMH.m 1.29 Method: spi_tr_mon Line: 6260 Multi-Coil T/R Driver Board Fault. An open circuit was detected on port 4. Check cable to multi-coil board or check for an open pin diode. 2225553 Multi-Coil T/R Driver Board Fault. An Open Circuit was Detected on Port 3. Thu Aug 21 12:06:10 1997 Suite: MRS1 Host: OC0 Proc: HPC_PORT_ID Error: 2225553 File: PMH.m 1.29 Method: spi_tr_mon Line: 6250 Multi-Coil T/R Driver Board Fault An open circuit was detected on port 3. Check cable to multi-coil board or check for an open pin diode.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 46-282460P1 STD, HEX BRASS UN INSULATED BANANA used on Multi coil plugs Banana Plugs guide pins used on Multicoil plugs 46-282460P1 Banana Plugs guide pins used on Multicoil plugs 46-282460P1
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.