Symptom
ID: 203884 : Fastcals fail to complete with bowtie length out of range
Error message
Error: 126169 Bowtie length out of range See long description for full error Mon Dec 15 11:43:02 Error: 126169 Suite: RP13 Host: SBC0 Proc: calMain File: AirCalJob.m 1.22 Method: jobDone Line: 813 Bowtie length out of range. Check collimator configuration. Bowtie length is : 1.05
Applies to
- Product
- HSA RP 1.X HSA RP 1.X with Zeus HSA Gantry and Sun Genesis Console FMI 25246 HSA/RP 1.6 Y2K Upgrade Special Tools Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-183027P128 REDUCED GAIN FILTER BRD.//T3202EK - used on CTi EDAS See Long Discription for Details Part #: 46-183027P128 Part Type: GP REPAIRABLES Part Weight: 1.36 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: REDUCED GAIN FILTER BRD.//T3202EK FRU: Yes Status: Orderable ICV/Transfer Price: 399.38 USD /Each List Price: 3594.42 USD /EACH
- Area
- CT