Pular para o conteúdo principal
GE Healthcare · CTLightSpeed

LightSpeed 2.X (Plus, QX/i) - Octane

Symptom

Software fails to boot with XIO Slot Reset Failure message on the screen. See the Long description for the full error code. Software fails to boot with XIO Slot Reset Failure message on the screen Error found in /usr/adm/SYSLOG Dec 14 11:44:52 6B:ct01 flash: 12/14/2005 16:42:36.000 GMT: XIO Slot Reset Failure *FAILED*

Error message

XIO Slot Reset Failure seen in /usr/adm/SYSLOG See Long Description for the full error code XIO Slot Reset Failure Type cd /usr/adm to see the boot logs Type more SYSLOG to see the errors. Dec 14 11:44:52 6B:ct01 flash: 12/14/2005 16:42:36.000 GMT: XIO Slot Reset Failure *FAILED*

Applies to

Models
LightSpeed
Product
LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2169940-36 replaces 2169940-3 Card 1 or 2 Graphics board used in the Octane computer part number 2169940-3 Card 1 or 2 Graphics board used in the Octane computer 2169940-36 replaces 2169940-3 Card 1 or 2 Graphics board used in the Octane computer Part #: 2169940-36 Part Type: GP REPAIRABLES Part Weight: 1.09 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: DUAL HEAD SE GRAPHICS OPTION/SPARE FRU: Yes Status: Orderable Comments: FLS - DUAL HEAD SE GRAPHICS OPTION /SPARE. U/O HELIOS & CTI 225MHZCOMPUTER. 1/22/99 MS08992. T3 CANDIDATE 2001. EOL QTR4 2001PCN 199401 2169940-36 REPLACES2169940-3. CANNOT ORDER MORE FROM SGI (PER JOHN MAYERS) SO SSS BLOCKED. HQT HISTORY: NEWEST TO OLDEST 030-1241-002 HU-DHSE GELS - THE -36 IS THE DUAL HEAD ON THE NEW OCTANE (225) FP 99/05,DUAL HEAD SOLD IMPACT GRAPHICS Part #: 2169940-3 Part Type: GP REPAIRABLES Part Weight: 1.81 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: OPTION DUAL HEAD GRAPHICS FRU: Yes Status: Orderable Comments: FLS - OPTION DUAL HEAD GRAPHICS USED ON NEXGEN COMPUTER MS08992 6/8/97 T3202KH PART HAS GONE END OF LIFE CONVERTING TO EXCHANGE. OUTRIGHT SALES MS08992 11/16/98 GWENN 199401 2169940-3 REPLACED BY 2169940-36 9/99 MS08992 2169940-3UA=J & M TRADING GELS - THE -3 HAS GONE END OF LIFE AND IS USED ON OLDER OCTANE 2169940-36 CAN BE DELIVERED INSTEAD. WAITING M.MEIER ABOUT THE LINKANBAN CSS F00 030825 IS REPLACED BY 2169940-36 ROUILES,DUAL HEAD
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required