CWA ID122714 Unable to Scan PROPELLER Combined with PURE due to Two Coils with Same Name.

Symptom

Unable to Scan PROPELLER Combined with PURE See Long Description iTrak # 13230090 The customer reports that their 8ch Head coil T2 PROPELLER sequence will not run. It was discovered that if PURE is turned ON with PROPELLER, the system will complete the auto-prescan but will not Prep-to-Scan or Scan. With PURE de-selected, the sequence will Scan. The same was found when using the GE protocol H.18 Propeller. If PURE is selected in this protocol, the system will complete the auto-prescan (or manual prescan) but will not Prep or Scan. The following errors were seen in the system error log: Fri Dec 12 07:40:21 2008 Error: 2258027 Host: lx-mr Process: LxMxConsumer File: DcPureImage.cpp Line: 158 Server Name: DP2 A Please perform another calibration scan. Fri Dec 12 07:40:22 2008 Error: 2254908 Host: lx-mr Process: LxMxConsumer File: gemrimxframework.appcontroller.scanner.ge Line: 0 Prep Failed. Please perform another calibration scan.

Error message

2254908 Prep Failed. Please perform another calibration scan. See Long Description Error: 2254908 Host: lx-mr Process: LxMxConsumer File: gemrimxframework.appcontroller.scanner.ge Line: 0 Prep Failed. Please perform another calibration scan.

2258027 Please perform another calibration scan. See Long Description Error: 2258027 Host: lx-mr Process: LxMxConsumer File: DcPureImage.cpp Line: 158 Server Name: DP2 A Please perform another calibration scan.

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR