Patient comfort sensor trip Due to Gradient Cooling Problem.

Symptom

Will Not Scan With Patient Comfort Sensor Trip. See Long Description Patient comfort sensor trip; cannot scan due to bore sensor indicating temperature is too hot for patient safety. A message "Patient Comfort Sensor Trip" appears in the message box on the left side of the scan page at the start of autoprescan.

Error message

2225684 Scan inhibited. Patient comfort sensor trip. See Long Description 2225684 ERROR: EM_SPI_BORE_TEMP_TRIP_2 Scan inhibited. Patient comfort sensor trip. To return to scanning, verify: 1. Patient fan is ON. 2. Scan room temperature is normal. 3. Air flow through the bore is unobstructed. 4. Gradient Cooling fan is normal. 5. BRM heat exchanger is normal.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2222564-30 50 and 60 HZ chiller by Lytron. See Long Description 2222564-30 50 and 60 HZ chiller by Lytron Model Number RC045J03BG0C011 Specified by 2222564PSP. Lower level FRU's should be utilized before attempting to replace the entire chiller. Gradient Chiller Motor - 2222564-25 Pump Assembly for Lytron Chiller - 2222564-29
Area
MR

Full cause & step-by-step resolution