GE Healthcare · CTLightSpeed230000036
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane
Symptom
"Scan hardware Stopped Scan" with scan data errors. "Scan hardware Stopped Scan" with scan data errors
Error message
Error 230000036 Active view data stopped with less data acquired than expected See long description for full error Wed Apr 23 21:20:26 2008 Error: 230000036 Host: ice Process: dataacq File: dc_actions.cxx Line: 2096 Function: Data Acquisition : Scan Data Acquisition Scan Type: Axial Scan: 9044/2/5 Scan Seq Id: 9000 Active view data stopped with less data acquired than expected. Only the acquired data will be saved. Expected active view data: 5e2f48h bytes Acquired active view data: 41418h bytes Missing active view data: 5a1b30h bytes Scan aborted.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2277527 H2 Gantry Fan Thermostat See Long Description for details Part #: 2277527 Part Type: GP CONSUMABLES Part Weight: 0.50 LBR Repairable: No Returnable: No Hazardous Material: N Description: GANTRY FAN THERMOSTAT FRU: Yes Status: Orderable Comments: FLS - H2 GANTRY: ORDER FROM MPE. MINIMUN ORDER QTY IS 25. MS26561 DSM 2/9/01
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.