GE Healthcare · CTLightSpeed18282518280450080
LightSpeed 2.X (Plus, QX/i) - Octane
Symptom
Scanner hardware reset and flash download fails
Error message
Error: 182825 HW/SW error: Communication Problems with the GANTRY DISPLAY. Open Long Description for full error Wed Dec 22 20:49:26 2004 Error: 182825 Host: ETC Process: Gantry Display File: gd_tnc_driver.c Line: 471 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 20:48:40:485 Log Series: 30 HW/SW error: Communication Problems with the GANTRY DISPLAY. Number of Communication Timeout errors: 0. Number of NACK's errors: 3. Number of DUART errors: 0. Numver of Checksum errors: 0. Error: 182804 The Gantry Display is NOT Initialized Properly See the LD for the full error code Error: 182804 The Gantry Display is NOT Initialized Properly Wed Feb 8 07:20:05 2006 Error: 182804 Host: ETC Process: Scan File: etc_in_mail.c 1.38 4/16/01 Line: 648 Function: No System Function Reported Scan: 50080/1/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 07:20:05:645 Log Series: 37 HW/SW error: Communication Problems with the Gantry Display. The Gantry Display is NOT Initialized Properly. (Scanning is NOT allowed if the Gantry Display is NOT Working)
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2256730-3 Display assembly used on 2.X/3X/4X5X/7X See long description for details. Part #: 2256730-3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: H-POWER COMPATIBLE GANTRY DISPLAY FRU: Yes Status: Orderable Comments: FLS - H-POWER COMPATIBLE GANTRY U/O CT H POWER DISPLAY
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.