GE Healthcare · CTLightSpeedVCT260134617
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Cradle does not move on GT table or stuck at the home position. Table moves OK, but Cradle will not drive IN or OUT of gantry.
Error message
Error : 260134617 Cradle clutch lines are disconnected. Cannot control cradle clutch See Long Description for complete error. Mon Nov 17 14:04:02 2008 Error: 260134617 Host: tgp Process: tgp_subsys File: gtcb_cmc_drv_req.cxx Line: 1040 Mon Nov 17 14:04:02 2008 Host : Table Sub-System Ermes # : 260134617 Exception Class : Secondary Severity : Pri/Soft File : gtcb_cmc_drv_req.cxx Line# : 1040 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Cradle clutch lines are disconnected. Cannot control cradle clutch
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5245271 GTCB3 (Global Table Control Board 3) This board replaces 5142177-3 GTCB2 and GTCB part # 5112470 thru 5112470-5, and 5142177. See long description for details Part #: 5245271 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: GTCB3 BOARD ASSY FRU: Yes Status: Orderable Comments: added per to ECO2080741 by sso#305010201 in May 8, 2009. Part will be available end of Sep, 2009, please order old p/n before new p/n on shelf.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.