Pular para o conteúdo principal
GE Healthcare · MRSigna22333962234203

2234203 QIN arbitration bus error Due to UCERD Digital board.

Symptom

Can Not Scan

Error message

2233396 The Frame Data Buffer Queue has overflowed. May need a longer TR time, or more time between echoes. Please restart scan. Thu Feb 22 17:59:07 2007 Error: 2233396 Host: MENEMR Process: NSP File: TPS:cap: c_int45 Line: 4 The Frame Data Buffer Queue has overflowed. May need a longer TR time, or more time between echoes. Please restart scan.

2234203 QIN arbitration bus error occurred. QIN is owned by device 0. QOUT is owned by device 0. Internal software error, please reset TPS. Error: 2234203 File: TPS:asc err isr Line: 501 A QIN arbitration bus error occurred. QIN is owned by device 0. QOUT is owned by device 0. Internal software error, please reset TPS.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
SecondOpinion Universal CERD Digital 2206099 See Long Description This part requires second opinion approval before ordering. See comments in resolution concept for second opinion requirements before ordering this part.