GE Healthcare · CTLightSpeedBrightSpeedHiSpeed23001708338587038101526011828830585

Scanner Hardware Stopped Scan

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 230017083 Possible beam obstructed - sin/cos vector magnitude = 0.xxxxxx - may include row and channel numbers See LD for complete error Mon Oct 31 07:08:03 2005 Error: 230017083 Host: cpcvct2p Process: accs File: aircal_spectral_acal_util.c Line: 445 Possible beam obstructed - sin/cos vector magnitude = 0.385870 Mon Mar 21 07:27:12 2011 Error: 230017083 Host: sj01_ct64s_6001 Process: accs File: aircal_spectral_acal_util.c Line: 449 Possible beam obstructed - row 58(26B), detector channel 375, sin/cos vector magnitude = 0.381015

Error: 260118288 (60-0332)too many kV_regul. number of kV_regul > number of KVRegulMax. (> 10 consecutives See LD for complete error Mon Sep 25 16:34:07 2006 Error: 260118288 Host: orp Process: orp_subsys File: gen_exec.c Line: 2376 Mon Sep 25 16:34:07 2006 Host : Jedi Ermes # : 260118288 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2376 Function : No System Function Reported Scan Type : Scout Scan: 30585/1/1 (60-0332)too many kV_regul. number of kV_regul > number of KVRegulMax. (> 10 consecutives (On Db configurable)) Z6,Z7= 14h,0h

Applies to

Models
LightSpeed, BrightSpeed, HiSpeed
Product
LightSpeed RT4 WB O G4 G HPT WB O M4 G HPT LightSpeed RT4 with GRE and MDAS4 or GDAS4 gantry At GE Medical Systems, we’re pioneering exciting CT technology for Radiation Therapy planning. Introducing the newest addition to our LightSpeed® series – LightSpeed RT, the first multi-slice, wide-bore CT scanner with the largest field of view in the industry. With a generous 80 cm gantry opening, the LightSpeed RT bore is 10 cm larger than conventional scanners. This provides complete flexibility for patient positioning freedom – even for the most demanding oncology needs. And the 65 cm variable field of view delivers 15 additional centimeters of visual coverage, allowing clinicians to see a patient’s entire anatomy for precise radiation therapy planning. Position and precision Thin, multi-slice imaging is the extra bonus – enabling respiratory gating, providing precise target delineation and delivering higher resolution digitally reconstructed radiographs (DRRs) than ever before. All adding up to the ability for you to deliver what really matters most – better cancer care for your patients. Plus, the LightSpeed RT scanner is available with the Exact™ couch insert, enabling you to consistently replicate patient immobilization for treatment. With LightSpeed RT, the future of your Radiation Oncology department is wide open. Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2266521-5 HV Tank (replaces 2266521-4 and 2266521-3 - backward compatable) used on LS 7.X, LS 5.X and LightSpeed RT. See Long Description for details 2266521-5 HV Tank (replaces 2266521-4 and 2266521-3 - backward compatable) used on LS 7.X, LS 5.X and LightSpeed RT. Dash Tanks includes the EMI Shielding needed for G-Das Part #: 2266521-5 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: JEDI H4 MPHVTANK WITH SHIELDS FRU: Yes Status: Orderable Part #: 2266521-4 Part Type: GP CONSUMABLES Part Weight: -5499945.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: JEDI H4 MPHVTANK WITH SHIELDS FRU: Yes Status: Orderable Part #: 2266521-3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 200.84 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: JEDI H4 MPHVTANK WITH SHIELDS FRU: Yes Status: Orderable Comments: added per ecr 2005924, as replacement for 2266521 and 266521-2. ms13452 040923 **If there is no Local Stock for the -3 part, you may order the -4 version which is backward compatible** ms0262 060815 *** 2ND OPINION NOTES *** Run the No-load KV test to verify HV tank is the error.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.