Pular para o conteúdo principal
GE Healthcare · CTLightSpeed18414217231184188170851713517140

LightSpeed 2.X (Plus, QX/i) - Octane

Symptom

Missed ISD See Long Description Inter Scan Delay ISD Missed .

Error message

Error: 184142 ISD Missed See Long description for complete error Wed Nov 8 11:01:40 2006 Error: 184142 Host: STC Process: Scan Control File: scTrigSeqActions.c 1.65 Line: 1043 Function: No System Function Reported Scan: 496/2/2 Type: Axial Exception Level: Pri/Soft Time: 11:01:40:295 Log Series: 937 ISD missed. Rx ISD: 1000 mS Requested ISD = 1 Second Actual ISD: 17231 Actual ISD = 17 Seconds More Data Follows: See Error: 184188 for additional detail

Error: 184188 Timing (in mS) since end of last scan Col:17085 DvRdy:17135 GenX:17135 TrigEn:17140 ExpEn:17230 See Long Description for more details Wed Nov 8 11:01:40 2006 Error: 184188 Host: STC Process: Scan Control File: scTrigSeqActions.c 1.65 Line: 1048 Function: No System Function Reported Scan: 496/2/2 Type: Axial Exception Level: Pri/Soft Time: 11:01:40:295 Log Series: 938 Last Hdw Rdys- Offsets:0x40 Cpb:0x20 Xray:0x80 Timing (in mS) since end of last scan. RxRcvd:-670 DoRdy:95 OfSt:165 Cpb:17085 OfStp:17095 Cr:810 Ax:100 GenP:175 Das:160 Col:17085 DvRdy:17135 GenX:17135 TrigEn:17140 ExpEn:17230 (Rdy:0xf00e2fff Of:0x540 Auto:0x57c Cpb:0x4057c Xray:0x80bff) This Error code is can give you a clue as to where the problem is comin

Applies to

Models
LightSpeed
Product
LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2169940-72UU (replaces 2169940-72) : Octane Computer 9 GB System Disk See Long Description for details Note: There are compatability problems with some consoles and disk drive firmware. Please see service note SN07112612a for details. Part #: 2169940-72UU Part Type: GP USED Part Weight: 2.43 LBR Repairable: No Returnable: No Hazardous Material: N Description: 9GB 10K RPM DISK WITH SLED FRU: Not Available Status: Orderable Comments: NOTE: MAY BE INACCURATE FOR H2 AND H3 CONSOLES. ** FE ASSIST ** 2169940-12 is no longer available can use 2169940-72 9GB or a 2169940-78 18GB Hard drives for a CTi or lightspeed QXI Octane systems as Image or OS disk. Will need to use updated LFC per 2282382-100 rev 2 or 2284291-100 rev 8. The 2169940-78 has 5000 images files where normal is 20000 if used as a CTi Image Disk. Refer to Service Note 07112612b. (212001261 8/11/09) ***OBSOLETE PART Case # 9746502*** This part has been identified as technically obsolete and no short term solution has been identified. FE must identify and order a higher or lower level spare as required. Contact your support engineering, On-line Engineering or Parts ID team should you require support. 212012173 7-24-236 Part #: 2169940-72 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 1.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: 9GB Octane 10K RPM DISK WITH SLED FRU: Yes Status: Orderable Parts Offered as: Current: 2169940-72 Used: Not Available Exchange: Not Available Comments: OCTANE 2 9GB DISK DRIVE USED ON HELIOS 2 AND 3 AND LX 9/25/03 CHECK WITH FIELD TO SEE IF THEY CAN USE THE 18GB PART# 2169940-78 IF NONE OF THE 9GB ARE AVAILIABLE
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.