Symptom

When DOWN elevation buttton is pressed the table continues to drive downward a little longer than normal - document in RFS - critical issue When DOWN elevation button is pressed the table continues to drive downward a little longer than normal.

Error message

Error: 260134707 Elevation could not stop with stop command. If table is moving downward, check valve See Long Description for error versions Mon Sep 19 08:02:03 2005 Error: 260134707 Host: tgp Process: tgp_subsys File: gtcb_emc_movechk.cxx Line: 616 Mon Sep 19 08:02:03 2005 Host : Table Sub-System Ermes # : 260134707 Exception Class : Secondary Severity : Pri/Soft File : gtcb_emc_movechk.cxx Line# : 616 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Elevation could not stop with stop command. If table is moving downward, check valve. last position = 793822 um current position = 794005 um commanded direction = UP time from stop = 587960 ms Mon Dec 12 10:22:24 2005 260134707 Host: tgp tgp_subsys gtcb_emc_movechk.cxx 579 Mon Dec 12 10:22:24 2005 Host : Table Sub-System Ermes # : 260134707 Exception Class : Secondary Severity : Pri/Soft File : gtcb_emc_movechk.cxx Line# : 579 Function : No System Function Reported Scan Type : None/Unknown Scan: 50189/1/0 Elevation could not stop with stop command. If table is moving downward, check valve. last position = 418827 um current position = 418382 um commanded direction = DOWN time from stop = 1040 ms

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5138908 TABLE SUPPORT TOOL JIG ASSY - used on Global tables See long description for details Part #: 5138908 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: TABLE SUPPORT JIG ASSY FRU: Yes Status: Orderable Comments: Not Available 5127284-3 (replaces 5127284) PROPORTIOAL VALVE ASSY Positoning GT - FRU yes - this assembly contains two proportional valves. See Long Description for details and picture: Part #: 5127284-3Part Type: GP CONSUMABLES Part Weight: 0.00 LBRRepairable: NoReturnable: NoHazardous Material: NDescription: PROPORTIOAL VALVE ASSY Positoning GTFRU: YesStatus: Orderable Comments: added per to ECO2079401 by sso#305010201 in Jun 8, 2009 Part #: 5127284 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: PROPORTIOAL VALVE ASSY Positoning GT FRU: Yes Status: Orderable Comments: Not Available This part is actually an assembly - the assembly contains two proportional valves and wire cable harness assembled/wired together. GT
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.