Pular para o conteúdo principal
GE Healthcare · MRSigna22227332225470

Will Not Scan With RF Errors due to SRFD2.

Symptom

Will Not Scan With RF Errors Will not scan with RF errors

Error message

2222733 MDS LINK ERROR The indicated peripheral is present but not functioning properly. It did not return a full data pkt in response to a read request. See Long Description Peripheral: %02xH (Gradient MDS = 0, RF MDS = 1) Reg addr: %02xH bytes exp: %d bytes rcvd: %d

2225470 RF Amplifier Response Timeout Fri Jun 11 09:39:08 2004 Error: 2225470 Host: Plymouth_MR Process: NSP File: HPC_PORT_ID:spi_cmb Line: 203 RF amplifier response timeout.^A RF MDS peripheral - RF Amp/RF Interface link failure. Verify that the RF Amplifier or RF Interface is powered on and that the RF MDS peripheral is connected to the RF Amplifier or RF Interface. Fri Jun 11 09:39:08 2004 Error: 2225470 Host: Plymouth_MR Process: NSP File: HPC_PORT_ID:spi_cmb Line: 203 RF amplifier response timeout. RF MDS peripheral - RF Amp/RF Interface link failure. Verify that the RF Amplifier or RF Interface is powered on and that the RF MDS peripheral is connected to the RF Amplifier or RF Interface.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573 See comments in Resolution Concept for second opinion requirements before ordering this part. SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues