Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT41827203982208200512101086910691944112

LightSpeed 3.X (Ultra, Plus, QX/i) - GRE

Symptom

Root Partition is almost full ( 98%) - customer is unaware of this condition and store log does not help. See Long Description for more info VCT Root Partition is almost full ( 98%) -Customer is unaware of this condition and store log does not help. Once the Root partion is full the System will reports file system is full to the customer. [root@qehct1]# df -k Filesystem 1K-blocks Used Available Use% Mounted on /dev/sdb2 4182720 3982208 200512 96% / /dev/sdb1 101086 4798 91069 6% /boot none 1944112 0 1944112 0% /dev/shm /dev/sdb9 9765248 272 9764976 1% /usr/interchange /dev/sdb10 11717120 792 11716328 1% /data /dev/sdb11 37670048 9633924 28036124 26% /usr/g /dev/md0 143363584 100004448 43359136 70% /usr/g/sdc_image_pool

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
06MW03.5 For Pathfinder 1.5 ! NOT FOR VCT ! VCT uses 06MW03.4 06MW03.5 For Pathfinder 1.5 ! NOT FOR VCT ! VCT uses 06MW03.4 06MW03.5 For Pathfinder 1.5 Reboot BEFORE and AFTER detailed cals. Used on 1X/2X/3X/4X/5X, BrightSpeed & HS QX/i with GRE Consoles ! NOT FOR VCT ! VCT uses 06MW03.4 Reboot BEFORE and AFTER detailed cals This software was released with FMI 25365 in 2nd quarter 2006. 06MW03C.5 Class C Software (tab 182 distribution) Part # 5165729 06MW03M.5 Class M Software (tab 182 distribution) Part # 5154540 Training for this software (including Network Filters) can be found at http://data.supportcentral.ge.com/upload/25823/doc_1221870.pdf BrightSpeed S/W Information: You may see 06MW03.4 or 06MW03.5 loaded on BrightSpeed Systems. Why two different Versions? The story is when the program team released the BrightSpeed M3 Pilot sites in FW10 2006, 06MW03.5 (PF1.5 M3 version) was not ready, so the program team decided to use 06MW03.4 as BrightSpeed M3 version. When 06MW03.5 was released, BrightSpeed systems phased 06MW03.5 into the production line. The only difference between the two versions is the "Rotor not at speed" issue which was resolved in 06MW03.5. For BrightSpeed systems running 06MW03.4, a patch is needed to resolve the "Rotor not at speed" issue. The patch was shipped with all BrightSpeed systems running 06MW03.4. There is no reason to upgraded BrightSpeed systems running 06MW03.4 with the rotor patch installed. For systems running 06MW03.4 verify the rotor patch is installed: [insite@ct01 ~]$ showprods | grep -i patch I patch 04/10/2006 The GNU patch command, for modifying/upgrading files. ver 2.5.4 rel 2.5.4.19 I MX200ROTOR_PATCH 04/10/2006 This is a MX200 Rotor patch for 06MW03.* ver 1 rel 1.1 For 06MW03.4 the correct classC & classM: ClassC = part# 5167665 Ver 06MW03C.4 ClassM = part# 5167597 Ver 06MW03M.4
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.