AMX700 The detector park switch was broken.
Definium AMX700 Unit drives in slow speed only due to a defective Detector Park Switch
Symptom
AMX700 The detector park switch was broken. Detector park switch also needs to be activated foe normal drive speeds. On this unit the detector park switch was broken. Jumpering out the detector park switch contact verified that the switch was defective and allowed maximum speed travel.
AMX700 AMX 700 drives in slow speed only. X-ray tube arm is parked and is sensed by the park switch hall effect sensor.
Applies to
- Models
- Definium
- Product
- Portables Definium AMX 700
- Area
- X-ray R&F
Full cause & step-by-step resolution