Proteus XRa Console KV and MA display blank, error 90-0703 - watchdog, asian interface board resetting due to damaged Co…

Proteus XRa Console KV and MA display blank, error 90-0703 - watchdog, asian interface board resetting due to damaged Collimator CAN BUS Cable.

Symptom

Proteus XRa Console KV and MA display blank, error 90-0703 - watchdog, asian interface board resetting.

Error message

Proteus XRa Error 90-0703 Proteus manual error code explanation: 90-0703H Watchdog reset has just occurred. This error is logged when the CPU of the kV control board is reset by it’s Watch Dog timer. 1. Software or Database problem. 2. kV control board failure

Applies to

Product
PROTEUS XRa
Component
Collimator Cable 2258383 This is the collimator CAN BUS cable used on Siemens and Eclipse Collimators on Proteus XRa systems
Area
X-ray R&F

Full cause & step-by-step resolution