GE Healthcare · X-ray R&F
Proteus XRa Console KV and MA display blank, error 90-0703 - watchdog, asian interface board resetting due to damaged Co…
Proteus XRa Console KV and MA display blank, error 90-0703 - watchdog, asian interface board resetting due to damaged Collimator CAN BUS Cable.
Symptom
Proteus XRa Console KV and MA display blank, error 90-0703 - watchdog, asian interface board resetting.
Error message
Proteus XRa Error 90-0703 Proteus manual error code explanation: 90-0703H Watchdog reset has just occurred. This error is logged when the CPU of the kV control board is reset by it’s Watch Dog timer. 1. Software or Database problem. 2. kV control board failure
Applies to
- Product
- PROTEUS XRa
- Component
- Collimator Cable 2258383 This is the collimator CAN BUS cable used on Siemens and Eclipse Collimators on Proteus XRa systems
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray R&F issues
- ADx-E Advantx-E console displayed a 6C00 Error (Digital Imaging Subsystem error - Error status received in IDC Prep Reply) when the first exposure of detector bad pixel calibration was attempted; cali…
- AEC Backup Time message
- New style ion chamber does not have a preamp
- Precision 500D Table Fluoro Carriage Front Bumper is not properly adjusted, allows excessive travel toward the front of the table.
- Precision 500D Aurora, dose readout does not update when an overhead Rad Tube Exposure is made, at any technic due to abnormally low mR/mAS from the Rad Overhead XRT.
- P500D Aurora with M4 Software (May 2006 Release), Dose Readout Correct on Console, Dose Printer Prints 0.0 due to a M4 Software Bug