Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2300065292508820000271220000271423000651010718

Lightspeed 1.X - Octane

Symptom

Images fail to display and recon shutdown. Images fail to display and recon shutdown

Error message

Error: 230006529 Recon Mgr is sending queue ack to queuemgr with a RECON_JOB_FAILED_RETRY status See Long Description for Full Error Tue Feb 1 06:33:34 2005 Error: 230006529 Host: ice Process: reconmgr File: rm_recon_events.c Line: 329 Function: Image Generation : SBC Processing Job Type: Prospective Axial Scan: 25088/2/9 Image: 25088/3/25 Recon Seq Id: 957 Recon Mgr is sending queue ack to queuemgr with a RECON_JOB_FAILED_RETRY status

Error: 200002712 MFM error: Format failure Group Number : (varies) Element : (varies) Error status : (varies)Error String : [ varies] See long description for full error. Error: 200002712 MFM error: Format failure Group Number : (varies) Element : (varies) Error status : (varies)Error String : [ varies]. The variable information gives the failing Group and Element and shows the failing string. Fri May 27 09:40:15 2005 Error: 200002712 Host: ct2 Process: imagecreate File: ic_mfm_main.c Line: 448 Image : 4243/1/1 Recon Sequence Id : 50 MFM error : Format failure Group Number : 40 Element : 254 Error status : -114 Error String : [ bad syntax for SH or LO or AT ]

Error: 200002714 MFM Error : Format failure DICOM image not created See Long description f

Applies to

Models
LightSpeed
Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.