GE Healthcare · MRSigna2244576
2244576 The GP reported SGA Under Voltage Due to SGA Power Suppy.
Symptom
Getting Undervoltage Faults on one Axis right after Bootup
Error message
2244576 The GP reported a %c Axis SGA UnderVoltage Fault 2244576 "The GP reported a %c Axis SGA UnderVoltage Fault.\n\ If no other faults reported, check cables or replace the\n\ SGA. Else, see the other fault error message."
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- SecondOpinion ACGD SGA Power Supply 2334862-2 This part requires second opinion approval before ordering. See comments in resolution concept for second opinion requirements before ordering this part.
- Area
- MR