LightSpeed 5.X (80/100kW)

Symptom

Dark ring and center circle artifacts seen in all images. Fastcal also fails to complete and reports Zslope errors. See long description for sample images. Dark ring and center circle artifacts seen in all images. Fastcal also fails to complete and reports Zslope errors.

Error message

Error: 230023041 Found the Zslope cal vectors are already 6 month old and need to update See the Long Description for the full error code Mon Oct 17 23:58:54 2005 Error: 230023041 Host: cts1 Process: dailyPrepRx File: DailyPrepUtils.c Line: 4394 Found the Zslope cal vectors are already 6 month old and need to update

Error: 230017124 Read minimum over maximum ratio from ZslopeCal.cfg is 0.550000. See LD for Full Error Error: 230017124 Read minimum over maximum ratio from ZslopeCal.cfg is 0.550000. Tue Dec 27 14:51:03 2005 Error: 230017124 Host: ct01 Process: accs File: zslope.c Line: 1910 Read minimum over maximum ratio from ZslopeCal.cfg is 0.550000.

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT