GE Healthcare · MR14603

OrthoMRI Incorrect magnet temperature.OrthoMRI Daily QA is failing.OrthoMRI Replaced the SCTERM unit.OrthoMRI Escan XQ.O…

OrthoMRI Incorrect magnet temperature.OrthoMRI Daily QA is failing.OrthoMRI Replaced the SCTERM unit.OrthoMRI Escan XQ.OrthoMRI 14603 N.

Symptom

OrthoMRI Daily QA is failing

Error message

OrthoMRI 14603 N: Signal/noise value isn't in the normal range

Applies to

Product
OrthoMRI Escan XQ EscanXQ and EscanNE
Software
OrthoMRI S/W Release 9.1B. OrthoMRI S/W Release 9.1B.
Component
OrthoMRI LNR9101148101 SCTERM Unit See long description. You must insert the correct SCTERM offset and the room temperature offset (read them from the label on the front side of the new SCTERM). Remember that the offsets are important to correctly read the magnet temperature errors but they don’t have any influence on the magnet warm up. The room temperature offset can be set only using step equal to ten units (e.g.: 10, 20, 30,……., 60, 70, etc)
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.