Application Software Very Slow due to Asterisks in Site Protocol Name.
Symptom
After Software Upgrade to 9.1 M4A FMI 60748 Protocol Download in Pelvis Subdirectory is VERY SLOW. Cannot Delete Corrupt Protocol. See Long Description The slow downloading of site protocols with strange characters in their names is a known problem with 9.1 M4A software. The workaround is to remove the strange characters from the protocol names or delete these protocols and recreate them using new names without the strange characters. iTrak 13206360 has been submitted. So the FE attempted to edit or delete the protocol in the "Pelvis" section that was causing the problem. This protocol had two asterisk symbols (**) in the beginning of the filename. But he was unable to edit it or delete it. When he tried to delete it, nothing happened. It did not get deleted. It would not delete.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR