GE Healthcare · CTLightSpeed2300110068350001139560721789815245047601000
LightSpeed 5.X (80/100kW)
Symptom
Scanning will not start when "START" button flashes
Error message
Error: 230011006 Unable to create directory on scan data disks. Requested scan can not be allocated. See long description for complete error Fri Feb 10 08:38:41 2006 Error: 230011006 Host: mct2 Process: scanmgr File: sdio_xfs_io_ops.c Line: 230 Function: Data Acquisition : SBC Processing Scan Type: Scout Scan: 0/0/0 Scan Seq Id: 835000 Unable to create directory on scan data disks. Requested scan can not be allocated. Scan: MCT2/1834/1/1 Directory name: /raw_data/ex1834.MCT2.1139560721.789815 Permissions: 775 (octal) Unix system error : 140 --- Remote I/O error. Error: 245047 Unable to allocate scan files See the long description for the full error code Thu May 12 17:43:37 2005 Error: 245047 Host: rrmcct16 Process: scanmgr File: scan_sequence.cxx Line: 1461 Function: System Control : Scan Control Scan Type: Scout Scan: 0/0/0 Scan Seq Id: 601000 Unable to allocate scan files Error: 245215 Scan setup failed on Step:3 (Allocate Scan File). See long description for full error Fri Sep 19 09:15:37 2008 Error: 245215 Host: ct10 Process: scanmgr File: scanmgr lib.cxx Line: 1167 Function: System Control : Scan Control S can Type: Scout Scan: 0/0/0 Scan Seq Id: 7000 Scan setup failed on St
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5114772 Darc2 Computer Dual Disk used ONLY on 16 slice MDAS and 4/8/16 slice GDAS See LD for more info 5114772 Darc2 Computer Dual Disk used ONLY on 16 slice MDAS and 4/8/16 slice GDAS Part #: 5114772 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: DARC Computer - Dual Disk H16 MDAS, GDAS FRU: Yes Comments: Added per GOC4 RSPL Replace the DARC2: 1) If you get a Westville DARC P/N 5114772 load the DARC software to recover. 2) If you get Jarrell p/n 5147442 perform LFC using 06MW29.7 software.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required