Pular para o conteúdo principal
GE Healthcare · Nuclear MedicineMillennium

MPR/MPS:Genieacq not booting to Login Screen hangs after message "Warning: Reset_card

Symptom

Genieacq not booting to Login Screen hangs after message "Warning: Reset_card : PCXE not found" Genieacq not booting to Login Screen hangs after message "Warning: Reset_card : PCXE not found". The message is normal on genieacq systems during bootup but normally continues to boot and display additional messages, problem is system is hanging at this point.

Error message

Warning - Reset_card: PCXE not found

Applies to

Models
Millennium
Product
MPR/MPS Millennium MPR Millennium MPS MG/MC/Myosight Millennium MG/MC Millennium MyoSIGHT IBACQ/X IBACQ/X IBACQ/X
Software
Genieacq All Genieacq All Genieacq All
Component
*Software Software Software
Area
Nuclear Medicine