2225536 Dynamic disable and T/R driver board error Due to Quad Extremity Coil.

Symptom

Can Not Scan

Error message

2225536 Dynamic disable and T/R driver board error. A spurious dynamic disable fault occurred. See Long Description Error: 2225536 Suite: MRS1 Host: OC0 Proc: HPC_PORT_ID PMH.m 1.29 Method: spi_tr_mon Line: 6220 Dynamic disable and T/R driver board error. A spurious dynamic disable fault occurred. It cleared itself before the specific problem could be read. Possible causes: noise on the dynamic disable board or a bad board.

Applies to

Models
Signa
Product
Signa Advantage Includes 4.x and 5.x systems. The software that this solution is tied to is indicated by the Software concept below.
Software
5.X
Component
46-328168G1 1.0T Quad Extremity Coil 46-328168G1 1.0T Quad Extremity Coil
Area
MR

Full cause & step-by-step resolution