GE Healthcare · CTLightSpeed26000312550304
LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry
Symptom
"Scan hardware Stopped Scan" with collimator errors. "Scan hardware Stopped Scan" with collimator errors
Error message
Error: 260003125 Did not detect a encoder index pulse when homing CAM B See LD for the complete error Error: 260003125 Did not detect a encoder index pulse when homing CAM B. Tue Sep 26 08:59:47 2006 Error: 260003125 Host: HSC Process: CCB Firmware File: aperture_home.c Line: 406 Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 50304/1/1 Type: Axial Exception Level: Pri/Soft Time: 08:59:47:285 Log Series: 37 Did not detect a encoder index pulse when homing CAM B.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5404254 (replaces 2125453-3) Collimator cam motor driver amp - used on 1.X (QXI) 2.X, 3.X, 4.X and 5.X. Not used on VCT See LD for details 5404254 2125453-3 Collimator cam motor driver amp -used on 1.X (QXI) 2.X, 3.X, and 4.X. Not used on VCT Part #: 5404254 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Helios Collimator CAM Drive FRU: Yes Status: Orderable Part #: 2125453-3 Part Type: GP CONSUMABLES Part Weight: 0.91 LBR Repairable: No Returnable: No Hazardous Material: N Description: HELIOS COLLIMATOR, CAM MOTOR DRIVER FRU: Yes Status: Orderable Comments: FLS - CAM MOTOR DRIVER HELIOS COLLIMATOR MS08992 9/3/98 M3 4TH QTR '98 HQT: VENDOR ORIENTAL MOTOR LEAD TIME WITH AIRFEIGHT 12 DAYS BY BOAT 6 WEEKS. INCREMENTAL COST FOR AIR FREIGHT $2
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.