GE Healthcare · MRSigna223845022384512238449
LX system won't scan due to 8645 Power Module PPBM errors on all 3 Gradient axis.
Symptom
Will Not Scan with Gradient Errors.
Error message
2238450 Y axis HIGH 8645 Power Module Poly Phase Buck Module Error occurred. Possible causes: high line volt. or overtemp. Tue Dec 21 11:24:59 2004 Error: 2238450 Host: MR1 Process: NSP File: HPC_PORT_ID:GAP/GIP/GP Line: 1030 Y axis HIGH 8645 Power Module Poly Phase Buck Module Error occurred. Possible causes: high line volt. or overtemp. Wait a few minutes for cooling, then turn 8645 power off for 30 seconds to clear error. After 30 seconds, turn 8645 power back on. Try to scan. If Main Control Board LED is lit again, replace PPBM. 2238451 Z axis HIGH 8645 Power Module Poly Phase Buck Module Error occurred. Possible causes: high line volt. or overtemp. Tue Dec 21 11:24:59 2004 Error: 2238451 Host: MR1 Process: NSP File:HPC_PORT_ID:GAP/GIP/GP Line: 1030 Z axis HIGH 8645 Power Module Poly Phase Buck Module Error occurred. Possible causes: high line volt. or overtemp. Wait a few minutes for cooling, then turn 8645 power off for 30 seconds to clear error. After 30 seconds, turn 8645 power back on. Try to scan. If Main Control Board LED is lit again, replace PPBM. 2238449 X axis HIGH 8645 Power Module Poly Phase Buck Module Error occurred. Possible causes: high line volt. or overtemp
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.