Pular para o conteúdo principal
GE Healthcare · MRSigna223845022384512238449

LX system won't scan due to 8645 Power Module PPBM errors on all 3 Gradient axis.

Symptom

Will Not Scan with Gradient Errors.

Error message

2238450 Y axis HIGH 8645 Power Module Poly Phase Buck Module Error occurred. Possible causes: high line volt. or overtemp. Tue Dec 21 11:24:59 2004 Error: 2238450 Host: MR1 Process: NSP File: HPC_PORT_ID:GAP/GIP/GP Line: 1030 Y axis HIGH 8645 Power Module Poly Phase Buck Module Error occurred. Possible causes: high line volt. or overtemp. Wait a few minutes for cooling, then turn 8645 power off for 30 seconds to clear error. After 30 seconds, turn 8645 power back on. Try to scan. If Main Control Board LED is lit again, replace PPBM.

2238451 Z axis HIGH 8645 Power Module Poly Phase Buck Module Error occurred. Possible causes: high line volt. or overtemp. Tue Dec 21 11:24:59 2004 Error: 2238451 Host: MR1 Process: NSP File:HPC_PORT_ID:GAP/GIP/GP Line: 1030 Z axis HIGH 8645 Power Module Poly Phase Buck Module Error occurred. Possible causes: high line volt. or overtemp. Wait a few minutes for cooling, then turn 8645 power off for 30 seconds to clear error. After 30 seconds, turn 8645 power back on. Try to scan. If Main Control Board LED is lit again, replace PPBM.

2238449 X axis HIGH 8645 Power Module Poly Phase Buck Module Error occurred. Possible causes: high line volt. or overtemp

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.