GE Healthcare · MRSigna2232902

2232902 Processor Exception Frame, Bus Error Due to UCERD Digital Board.

Symptom

Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis

Error message

2232902 Processor Exception Frame, Bus Error: vector 2, interrupt level 0 Mon Nov 3 13:11:30 1997 Error: 2232902 Host: oc001 Process: NSP File: TPS:excCHECKER Line: 0 Processor Exception Frame, Bus Error: vector 2, interrupt level 0 current pc 0x279bf2, sr 0x3004, task name MC860_CEID_2, fault address 0x400b0008, fault status 0x105 additional information in /usr/g/service/log/vxExLg-4c752740

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
SecondOpinion Universal CERD Digital 2206099 See Long Description This part requires second opinion approval before ordering. See comments in resolution concept for second opinion requirements before ordering this part.
Area
MR

Full cause & step-by-step resolution