GE Healthcare · MRSigna2225686

2225686 The bore temperature reading is low due to Lytron chiller set point too low.

Symptom

Getting a message that the Bore Temperature is too low. Pop-up message stating that the bore temperature is too cool stays on the screen briefly, then goes away. Scanning is not paused.

Error message

2225686 The bore temperature reading is low. Error: 2225686 Host: LXMR Process: NSP File: HPC PORT ID:spi bore wall temp mon Line: 400 The bore temperature reading is low. Check for low room temperature, a short in the sensor, sensor cable, or connector, or a problem with the piggy-back amplifier board in the SRI.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution