GE Healthcare · CTLightSpeed245207
LightSpeed 4.X - GRE
Symptom
Unable to scan with a message on screen that says: Collimator/DAS reset failed.
Error message
Error: 245207 Software has detected an invalid or mistached DAS type. This will prevent the system from scanning. see long description Mon Apr 3 08:29:12 2006 Error: 245207 ls16 scanmgr validate request.cxx 378 Software has detected an invalid or mistached DAS type. This will prevent the system from scanning.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5191271 MDAS 16 Fuse Holder Upgrade Kit. Supercedes part number 46-170042P28. See Long Description for more information. Part #: 5191271 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: MDAS 16 Fuse Holder Upgrade Kit FRU: Yes Status: Orderable Comments: Per ECR 2036788, 212008898, 11/30/2006. MDAS Fuse Holder Replacement Kit Contents: Part #: 46-170042P28 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: 30A, 250V PANEL-MOUNT FUSEHOLDER. FRU: Yes Status: Orderable Comments: FLS - FUSE HOLDER PART OF FMI KIT ** FE Assist ** IF FUSE HOLDER P/N 46-170042P28 FAILED, PLEASE ORDER MDAS FUSE FIX ON FAILURE KIT 5191271 (12/12/06 212001261)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.