GE Healthcare · Nuclear MedicineMillennium
Genieacq Dynamic acquisition display stops updating after about 15 minutes into scan.
Symptom
Genieacq Dynamic acquisition display stops updating after about 15 minutes into scan. Genieacq Dynamic acquisition display stops updating after about 15 minutes into scan Image display appears frozen and does not change but frames increament, if allow scan to complete and store the data shows that scan did acquire normally and frames were updating. System running 4.0 software Dynamic scan with 256 matrix. ELGem34164: CQA 13046937 MILL MPR Display not updating after 15 min. on a long dynamic acq.
Error message
fep "Frame 0 requested, last frame is 1" followed by fep "Request framing board to Flush frame set for det 1, es 0:1, br 0, frm 0 to 0" 2008/07/30 14:45:13 fep "Frame 0 requested, last frame is 1"2008/07/30 14:45:13 fep "Request framing board to Flush frame set for det 1, es 0:1, br 0, frm 0 to 0" This shows that display's request for the latest acquired frame (frame 0) failed. The specific error code is not logged, but the request failed at the framing library level in a call to frm_lib.c:frm_read_acq_data(). The errors continue; 2008/07/30 14:45:19 fep "Frame 0 requested, last frame is 2" 2008/07/30 14:45:19 fep "Request framing board to Flush frame set for det 2, es 0:1, br 0, frm 0 to 0" ... 2008/07/30 14:45:19 fep "Frame 0 requested, last frame is 2" 2008/07/30 14:45:19 fep "Request framing board to Flush frame set for det 1, es 0:1, br 0, frm 0 to 0" ... 2008/07/30 14:45:53 fep "Frame 0 requested, last frame is 8" 2008/07/30 14:45:53 fep "Request framing board to Flush frame set for det 2, es 0:1, br 0, frm 0 to 0" Information Source Tim Jones
Applies to
- Models
- Millennium
- Product
- MPR/MPS Millennium MPR Millennium MPS MG/MC/Myosight Millennium MG/MC Millennium MyoSIGHT
- Software
- Genieacq 4.0 Genieacq 4.0
- Component
- *Software Software Software
- Area
- Nuclear Medicine
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare Nuclear Medicine issues
- How do I capture and print COR and Camera Uniformity (CUA) result screens on Genieacq, then send to a processing station for ACR
- Collision and E-stop lights are red on the back of the gantry and no movement with Hand Held Controller (HHC) or the genieacq.
- Need part numbers for the Millennium MG/MC table E-Stop
- How do I Delete a file on Genieacq that the system thinks is still acquiring
- VG, the contour works... but in the opposite way.
- MG_MC_MyoSight_MPR_MPS Bar Phantom Resolution Size