GE Healthcare · MRSigna2238501
TPS reset failure and error 2238501 caused by Defective Gradient Power Module.
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2238501 The Y Axis LOW 8645 Power Module is reporting all its faults to the MIF. 2238501 Host: lx Process: NSP File: HPC PORT ID:GAP/GIP/GP Line: 1060 Check Y Axis LOW 8645 Power Module Power.^A The Y Axis LOW 8645 Power Module is reporting all its faults to the MIF. This is most likely caused by power loss to the power module.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 46-328206P115 CROWN A10999-G2 8645 PWR MOD W/CROWB
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.