GE Healthcare · CTLightSpeed26010002850002
LightSpeed 5.X (80/100kW)
Symptom
After pressing the “Confirm” button in the patient scan interface, the “Start Scan” key blinked for at least one minute then axial or helical scan was paused. After pressed the “Confirm” button in the patient scan interface, wait the “Start Scan” key blinked for at least one minute and axial or helical scan was paused.
Error message
Error: 260100028 Scan Control timed out waiting for the Ready For Views message from the OC. See Long Description for complete error. Thu Apr 13 22:25:49 2006 Error: 260100028 Host: tgp Process: tgp_subsys File: scTrigSeqActions.c Line: 1616 Thu Apr 13 22:25:50 2006 Host : Table Gantry Processor Ermes # : 260100028 Exception Class : Abort Severity : Primary File : scTrigSeqActions.c Line# : 1616 Function : DATA ACQUISITION : Scan Data Acquisition Scan Type : Axial Scan: 50002/1/48 Scan Control timed out waiting for the Ready For Views message from the OC. This message is sent from the OC once offset collection has completed. Possible causes: trigger failure, Data Collection failed to receive data, software error. TMO: 5000 msec .
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5139590 Control assembly left side gantry. ( replaces part# 2256729 and 2330695 ) – used on H2 & HP gantry’s (LightSpeed 2.x, 3.x, 4.x, 5.x, 7.x ) FEMC info in Long Description. 5139590 Control assembly left side gantry. ( replaces part# 2256729 and 2330695 ) – used on H2 gantry’s (LightSpeed 2.x, 3.x, 4.x, 5.x, 7.x ) Part #: 5139590 Part Type: GP CONSUMABLES Part Weight: 4.41 LBR Repairable: No Returnable: No Hazardous Material: N Description: H POWER OPERATOR CONTROLS LEFT SIDE FRU: Yes Status: Orderable If part number 2330695 is ordered it will automaticaly cross to part 5139590 Old part info 2256729 Control assembly left gantry Part #: 2256729 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: CONTROL ASM LEFT H2 GANTRY FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.