Millennium MPR/MPS Unable to characterize roll axis, fails to calibrate Roll axis error # 15 During calibratio…

Millennium MPR/MPS Unable to characterize roll axis, fails to calibrate Roll axis error # 15

Symptom

Millennium MPR/MPS Unable to characterize roll axis, fails to calibrate Roll axis error # 15 During calibration roll axis will move the approx -135 degrees and stop with an error message "Calibration failed on Roll axis - error #15"

Error message

Calibration failed on Roll axis - error #15 Error log entry when it fails: 2008/07/30 23:54:56 EST genieacq:gip_ph;83:gip_recv.c[$Revision: 1.63.2.16 $] handleCntlMsg() line 1301 error 0x07100005 level 2 "Calibration error reported on the Roll axis." 2008/07/30 23:54:56 EST genieacq:tgc_ph;102:moveio.c[$Revision: 1.11.2.6 $] mio_gip_r_cal() line 1624 error 0x14300019 level 3 "GIP_R_CALIBRATE: Error 15 for ROLL axis." 2008/07/30 23:54:56 EST genieacq:tgc_ph;102:calib.c[$Revision: 1.37.2.3 $] cal_gip_cal_reply_proc() line 446 error 0x14300003 level 2 "ROLL calibration failure."

Applies to

Models
Millennium
Product
MPR/MPS Millennium MPR Millennium MPS
Component
*Gantry Gantry Gantry 2227404 Proximity Sensor PHOTO EYE for ROLL DRIVE ASSEMBLY (Old P/N 2157085) 2227404 Proximity Sensor PHOTO EYE for ROLL DRIVE ASSEMBLY (Old P/N 2157085) Part #: 2227404 Part Type: GP CONSUMABLES Part Weight: 0.07 LBR Repairable: No Returnable: No Hazardous Material: N Description: PROXIMITY SENSOR FRU: Yes Status: Orderable
Area
Nuclear Medicine

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.