LightSpeed 4.X - GRE
Symptom
Scanned images display in ExamRx viewports but not in Image works and prospective recons failed with image installation errors. See Long Description for details Scanned images display in ExamRx viewports but not in Image works and prospective recons failed with image installation errors. When attempting to call up images in ImageWorks viewer the entire scan series does not even list when listing out the exam. Recon status on right side display monitor will indicate failures even though images display in the autoview viewport.
Error message
From the var/adm more messages -> SCSI Error: (1:2:0) Status=02h (CHECK CONDITION) - Key=3h (MEDIUM ERROR); FRU=D2h - See LD for more information SCSI errors point to scsi id 2 in /var/adm/messages Oct 17 07:10:23 bhmgect4 kernel: SCSI Error: (1:2:0) Status=02h (CHECK CONDITION) Oct 17 07:10:23 bhmgect4 kernel: Key=3h (MEDIUM ERROR) ; FRU=D2h Oct 17 07:10:23 bhmgect4 kernel: ASC/ASCQ=16h/00h "" Oct 17 07:10:23 bhmgect4 kernel: CDB: 28 00 00 88 B8 3F 00 00 08 00 Oct 17 07:10:23 bhmgect4 kernel: Oct 17 07:10:29 bhmgect4 kernel: SCSI Error: (1:2:0) Status=02h (CHECK CONDITION) Oct 17 07:10:29 bhmgect4 kernel: Key=3h (MEDIUM ERROR); FRU=D2h Oct 17 07:10:29 bhmgect4 kernel: ASC/ASCQ=16h/00h "" Oct 17 07:10:29 bhmgect4 kernel: CDB: 28 00 00 88 B8 3F 00 00 08 00 Oct 17 07:10:30 bhmgect4 kernel: scsi3: ERROR on channel 0, id 2, lun 0, CDB: Read (10) 00 00 88 b8 3f 00 Oct 17 07:10:30 bhmgect4 kernel: Info fld=0x88b840, Current sdd: sense key Medium Error Oct 17 07:10:30 bhmgect4 kernel: Additional sense: Data synchronization mark error Oct 17 07:10:30 bhmgect4 kernel: end_request: I/O error, dev sdd, sector 8960064 Oct 17 07:10:30 bhmgect4 kernel: I/O error in filesystem ("md0") meta-data dApplies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required