Symptom

Image recon is failing intermittently - document in RFS - critical issue See long description for details. Image recon is failing intermittently.

Error message

Error: 200110026 OC Processing Can not find scout scans See long description for full error Sun Dec 23 16:25:29 2007 Error: 200110026 Host: vct1 Process: scanRx File: ViewEdit.cxx Line: 9807 Function: Data Acquisition : OC Processing Can not find scout scans. AutoMa will be disabled for Series:3

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5147443-2 Jarrell IG for NON VCT systems (replaces 5147443 ). See long description for details Part #: 5147443-2 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Jarrell GRE IG with RAC rev.D or Higher FRU: Yes Status: Orderable Comments: Per ECR 2040301 (Part of Jarrell NPI), 212008898, 3/8/07. **FE Assist** applies to CT but not PET systems. Must order software release 07MW11.10, p/n 5220443 (deployed by FMI 25389), and perform a LFC to support this FRU if system is not currently running sw rls 06MW29.7 or 07MW11.10 (8/16/07 212001261) 212008041 12/03/2007 15:23 GMT Part #: 5147443 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Computer - Image Generation, 2nd Generation FRU: Yes Status: Orderable Comments: Per Jarrell RSPL, 7/19/06, ms11892. ** FE Assist ** If ordering this Jarrell IG to replace a Westville IG 2362872 must also order a 5127222-2 Shielded Network cable and replace the existing one on your system Plus, must order software release 06MW29.7, p/n Part #: 5193989 and perform a LFC. (12/8/05 212001261
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.