Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26011631950132260115244750001703000101663

LightSpeed 5.X (60kW)

Symptom

Table would not move in or out. Table would not move in or out

Error message

Error: 260116319 State Response Timeout Occurred See Long Description for full error Fri Sep 2 07:22:58 2005 Error: 260116319 Host: tgp Process: tgp_subsys File: subsys_mgr.c Line: 2343 Fri Sep 2 07:22:55 2005 Host : Table Gantry Processor Ermes # : 260116319 Exception Class : Secondary Severity : Pri/Soft File : subsys_mgr.c Line# : 2343 Function : No System Function Reported Scan Type : None/Unknown Scan: 50132/1/0 State Response Timeout Occurred Subsystem: Table Gantry Processor State: 12 Non-Responding App Id: 100 Timeout Value: 1000 msec.

Error: 260115244 MTCB cradle moving direction error See the Long Description for the complete error code Thu Dec 28 10:59:18 2006 Error: 260115244 Host: tgp Process: tgp subsys File: mtcb cradle monitor.c Line: 823 Thu Dec 28 10:59:18 2006 Host : Table Sub-System Ermes # : 260115244 Exception Class : Secondary Severity : Pri/Soft File : mtcb cradle monitor.c Line# : 823 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 MTCB cradle moving direction error. Possible Encoder, cabling or MTCB board issue. Current Speed = -7300 Target Speed = 75000 Target Position = 1703000 Current Position = 101663

Error: 260111023 The 

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5263397 MTCB board used on lightspeed 5.X See 2ND OPINION NOTES in Long Description for details 5263397 MTCB board used on lightspeed 5.X Part #: 5263397 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 KGM Repairable: No Returnable: Yes Hazardous Material: N Description: Assembly-MTCB-3 mounting panel, Hpower-table FRU: Not Available Status: Orderable Comments: Per ECR 2050004, 212008898, 11/06/07. *** 2ND OPINION NOTES *** 5263397 can be USED on PET systems only. CT systems should order 5120650 unless stock is depleted, then 5263397 is acceptable. 111607 212010493
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues