GE Healthcare · CTLightSpeed178575
LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2
Symptom
System Shutdown and Reboot during OC Initial See Long Description for Detail System Shutdown and Reboot during OC initial Power on the console, system performing “ OC initialization”, soon after system shutdown automatically, when the “system power off OK” message appear , system startup again automatically. Clatter noisy from the scanbox, and a LED blinking on the RIP board and synchronize the phase of the sound from the scan box.
Error message
Error: 178575 Failed to R-HARD the following controllers: STC OBC ETC. See long description for full error Fri Oct 8 07:49:44 2004 178575 2 ct03 fwmgr HwSetUp.c 837 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failed to R-HARD the following controllers: STC OBC ETC.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2 This is HP CPU host running LINUX with Pegasus IG. 308L.8_H3.3M4 ~ 4 or 8 slice MDAS. ~ Could be an Ultra, Plus or QXi system. Must run swokinstall -p to verify. ~ See Long description for details . LightSpeed 3.X MDAS4 Gantry and GOC2 Gantry - 2281177-2, DAS - 2272768, Console – 2341104-2, Table - 2276776, PDU - 2269902 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2260421 replaces 2260421-2 VME console power supply. Used on CTi - 4.x H3 GOC1 and GOC2 Operators Console. See Long Description for Details. 2260421 replaces 2260421-2 Part #: 2260421 Part Type: GP REPAIRABLES Part Weight: 2.18 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: VME CONSOLE POWER SUPPLY ASSEMBLY FRU: Yes Status: Orderable Part number 2260421-2 is a comsumable assembly at EOL with 45 per year comsumsion. Part number 2260421 is the same base assembly with an added bracket and is a repairable. Do to the EOL issue and low comsumption of 45 per year we will eliminate p/n 2260421-2 and replace it with p/n 2260421 to have one repairable part. This will eliminate the need to stock two parts and create a new repair process.
- Area
- CT