Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT2900120122900090021183997163967569200352903

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Reconstruction Error detected during boot-up - reconSelftest. Reconstruction Error detected during boot-up - reconSelftest . Pop ping pop up box comes up. Attention: Reconstruction Process has stopped.

Error message

Error: 290012012 Error from VRAC See Long Description for full error Fri Apr 3 15:19:31 2009 Error: 290012012 Host: ig3 Process: image_generation File: VRACInterface.cpp Line: 483 Error from VRAC

Error: 290009002 Error failed in opening file See long description for full error Mon Jul 9 17:34:45 2007 Error: 290009002 Host: darc Process: recon_control File: IsrUpdateCache.cxx Line: 153 Error failed in opening file. file name = /usr/g/queue/priority3_jobs/FCCC.3776.2.1177.1183997163.967569.pro_cine error msg = No such file or directory

Error: 200352903 Reconstruction errors detected See Long Description for error. Fri Jul 29 20:27:00 2005 Error: 200352903 Host: whct Process: reconSelftest File: reconSelftest.c Line: 200 Reconstruction errors detected Had Recon Retry on image 414/4/1, the recon job did not succeed errors in 1 images during selftest --------------------------------------------------------------------------- Fri Nov 17 12:32:05 2006 Error: 200352903 Host: ct04 Process: reconSelftest File: reconSelftest.c Line: 201 Reconstruction errors detected Had [IG 1] image checksum mismatch errors in 4 images during selftest

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5114533-2 (Westville) VIG for VCT console. See Long Description for details. 5114533-2 (Westville) VIG for VCT Part #: 5114533-2 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: VIG Computer - with updated VRAC FRU: Yes Status: Orderable Comments: added per ECR: 2025620. This part is replacement for 5114533. Do not use a 5114533 to replace a 5114533-2 . ms13452 060221 ** FE Assist ** If no stock of Westville VIG will need to order a Jarrell 5159834-2 and order software 29.7 or later pn 5193989 Refer to Service Note 06101010A. *** 2ND OPINION NOTES *** Bad software has looked like a hardware failure. Try darc load first. Beeps from IG indicate bad hardware. Also RJ45 cables can cause intermittent failures.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required