GE Healthcare · CT2000027182639020000043720000042323000904040532
Lightspeed 1.X - Octane
Symptom
Images fail to display and recon shutdown. Images fail to display and recon shutdown
Error message
Error: 200002718 Image create Pixel data request failed See the long description for the full error code Thu Jun 30 23:03:01 2005 Error: 200002718 Host: ip01 Process: imagecreate File: ic imagecreate req.c Line: 247 Image : 26390/2/3 Recon Sequence Id : 15 Image create Pixel data request failed Error: 200000437 Image installation failed for image (for some Exam/Series/Image number) Open Long Description for full error varaiations Sat Jul 16 13:11:02 2005 Error: 200000437 Host: hsa2 Process: dms File: InstallImage.cxx Line: 316 Image installation failed for image 7297/1/1. Reason: Response from SdC timeout. Thu Feb 2 13:32:10 2006 Error: 200000437 Host: ct01 Process: dms File: InstallImage.cxx Line: 713 Image installation failed for image 2336/999/1. Reason: DMS internal error occurred. image path = %s Error code: 200000423 Error: 230009040 The image did not install in the database within the time allotted. See LD for an error example. Tue Sep 27 09:32:40 2005 Error: 230009040 Host: ct04_sbc Process: queuemgr File: qmgr_queue_cntl.c Line: 1773 Function: Image Generation : SBC Processing Job Type: Prospective Helical Scan: 40532/2/1 Image: 40532/2/20 The image did not install in th
Applies to
- Product
- Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 10.5_2.8.2I_H1.3M4 for LightSpeed 1.X (QXi) with Console Assembly (2180551) (Console with O2 Computer) or (2180551-2) (Console without O2 Computer) - showprods returns LightSpeed Apps ver 10.5_2.8.2I_H1.3M4 Released via FMI 25269 The features introduced with this software include(but are not limited to) the following:Thin Twin Axial Mode (2 x .63 mm) Multi-Destination DICOM Print Capability Latest release of Iniste Interactive (IIP 2.1) Advanced Artifact Reduction (AAR) Web Browser Based Common Service Desktop - Phase II Includes a new HomePage New Log Viewer Separated Class A, C, and M software Beam-Tracking Improvement LightSpeed Apps ver 10.5_2.8.2I_H1.3M4 Released via FMI 25269 The features introduced with this software include (but are not limited to) the following: Thin Twin Axial Mode (2 x .63 mm) Multi-Destination DICOM Print Capability Latest release of Iniste Interactive (IIP 2.1) Advanced Artifact Reduction (AAR) Web Browser Based Common Service Desktop - Phase II Includes a new HomePage New Log Viewer Separated Class A, C, and M software Beam-Tracking Improvement LightSpeed QX/i Rev 1.3 System Software Collector 2260334-5 consisting of 1.3 Software CD-ROM Collector 2258825-7 and Software Rating Plate 2240213-2 1.000 PC QX/I IRIX 65 VER 6.8.A DI 2240215-2 1.000 PC QX/I IRIX 65 VER 6.8.A DI 2240216-2 1.000 PC QX/I IRIX 65U VER 6U4.5.A 2258610-6 1.000 PC LIGHTSPEED QX/I APPS SFTW
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required