GE Healthcare · CTHiSpeed2300065073392823000904123000906710989200002748
HiSpeed QX/i - Octane
Symptom
Image reconstruction failures. Image reconstruction failures
Error message
Error: 230006507 Reconmgr timed out waiting for IG done from IG Control See long description for a full sample error. Tue Aug 19 12:10:21 2008 Error: 230006507 Host: ice Process: reconmgr File: rm process job.c Line: 817 Function: Image Generation : SBC Processing Job Type: Prospective Helical Scan: 33928/2/1 Image: 33928/2/84 Recon Seq Id: 991 Reconmgr timed out waiting for IG done from IG Control SDC Setup:102 SDC Done:0. IG Setup:200 IG Done:0. Error: 230009041 The image did not complete reconstruction within the time allotted. See long description for complete error Tue Apr 15 19:45:47 2003 Error: 230009041 Host: smopct1 Process: queuemgr File: qmgr_queue_cntl.c Line: 2524 Function: Image Generation : SBC Processing Job Type: Retro Axial Scan: 414/4/1 Image: 414/4/5 The image did not complete reconstruction within the time allotted. Error: 230009067 Retry queue entry See Long Description for error example. Tue Oct 19 09:24:04 2004 Error: 230009067 Host: dgct Process: queuemgr File: qmgr_list_mgmt.c Line: 2217 Function: Image Generation : SBC Processing Job Type: Prospective Helical Scan: 10989/2/1 Image: 10989/2/1 Retry queue entry Error: 200002748 Timeout waiting for pixel
Applies to
- Models
- HiSpeed
- Product
- HiSpeed QX/i - Octane Always MDAS4 101I.2_HSQXIM5 101.4_HSQXIM4 HiSpeed QX/i - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 101I.2_HSQXIM5 101.4_HSQXIM4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 101I.2_HSQXIM5 See LD for ClassM & C 101I.2_HSQXIM5 used on HiSpeed (YMS) only Class M software - 2395173 or 2401759 (6 pack) Class C software - 2395172 or 2401759 (6 pack) Patch G.6 Mandatory FMI# 25363 (P/N 5136941)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required